User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: RCA "Senior VoltOhmyst" (VOM) Model WV-98C DATE: 15-May-95 AUTHOR: User Contributed REVISION: A ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 46 NUMBER OF LINES: 243 CONFIGURATION: Fluke 5100B CONFIGURATION: HP 3458A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # RCA "Senior VoltOhmyst" (VOM) Model WV-98C: Cal DC/AC and Ohms ONLY. # Resistance Tolerance ń10% of reading. (Unknown reference spec.) 1.001 RSLT =Accuracy: Manufacturers specification except for the 1.002 RSLT = following: Ohms ń10% of nominal. 1.003 RSLT = Calibrated DC/AC and Ohms functions ONLY. 1.004 RSLT =Reference Documents: Manufacturers manual. 1.005 RSLT = 1.006 ASK+ R X 1.007 ASK- F 1.008 DISP Allow the UUT to stabilize at ambient 1.008 DISP conditions for 30 minutes. 1.009 HEAD ELECTRICAL ZERO 1.010 DISP Set the FUNCTION switch to +DC VOLTS 1.011 DISP Set the PROBE switch to DC 1.012 DISP Set the RANGE switch to .5V DC 1.013 DISP Adjust the "ZERO ADJUST" control for a reading of 1.013 DISP 0 VDC. 1.014 HEAD {+DC Volts Calibration} 1.015 DISP Connect the probe tip to the 5100B "OUTPUT HI". 1.015 DISP Connect the ground clip to "OUTPUT LO" 1.016 5100 0.5 0.400V 0.015U 2W #! Test Tol 0.015, Sys Tol 4.1e-005, TUR 365.8537 (>= 4.00). 2.001 DISP Re-Zero the meter after changing to the 1.5V range. 2.002 5100 1.5 1.400V 0.045U 2W #! Test Tol 0.045, Sys Tol 8.1e-005, TUR 555.5556 (>= 4.00). 3.001 5100 5 4.000V 0.15U 2W #! Test Tol 0.15, Sys Tol 0.000365, TUR 410.9589 (>= 4.00). 4.001 5100 15 14.000V 0.45U 2W #! Test Tol 0.45, Sys Tol 0.000765, TUR 588.2353 (>= 4.00). 5.001 5100 50 10.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.000605, TUR 2479.3388 (>= 4.00). 6.001 5100 50 20.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.0028, TUR 535.7143 (>= 4.00). 7.001 5100 50 30.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.0032, TUR 468.7500 (>= 4.00). 8.001 5100 50 40.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.0036, TUR 416.6667 (>= 4.00). 9.001 5100 50 50.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.004, TUR 375.0000 (>= 4.00). 10.001 5100 150 140.00V 4.5U 2W #! Test Tol 4.5, Sys Tol 0.0076, TUR 592.1053 (>= 4.00). 11.001 5100 500 400.00V 15U 2W #! Test Tol 15, Sys Tol 0.027, TUR 555.5556 (>= 4.00). 12.001 5100 1500 900.00V 45U 2W #! Test Tol 45, Sys Tol 0.047, TUR 957.4468 (>= 4.00). 13.001 HEAD {-DC Volts Calibration} 13.002 DISP Set the FUNCTION select switch to -DC VOLTS 13.003 DISP Re-Zero the meter for the .5V DC range 13.004 5100 0.5 -0.400V 0.015U 2W #! Test Tol 0.015, Sys Tol 4.1e-005, TUR 365.8537 (>= 4.00). 14.001 DISP Re-Zero the meter after changing to the 1.5V range. 14.002 5100 1.5 -1.400V 0.045U 2W #! Test Tol 0.045, Sys Tol 8.1e-005, TUR 555.5556 (>= 4.00). 15.001 5100 5 -4.000V 0.15U 2W #! Test Tol 0.15, Sys Tol 0.000365, TUR 410.9589 (>= 4.00). 16.001 5100 15 -14.000V 0.45U 2W #! Test Tol 0.45, Sys Tol 0.000765, TUR 588.2353 (>= 4.00). 17.001 5100 50 -10.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.000605, TUR 2479.3388 (>= 4.00). 18.001 5100 50 -20.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.0028, TUR 535.7143 (>= 4.00). 19.001 5100 50 -30.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.0032, TUR 468.7500 (>= 4.00). 20.001 5100 50 -40.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.0036, TUR 416.6667 (>= 4.00). 21.001 5100 50 -50.000V 1.50U 2W #! Test Tol 1.5, Sys Tol 0.004, TUR 375.0000 (>= 4.00). 22.001 5100 150 -140.00V 4.5U 2W #! Test Tol 4.5, Sys Tol 0.0076, TUR 592.1053 (>= 4.00). 23.001 5100 500 -400.00V 15U 2W #! Test Tol 15, Sys Tol 0.027, TUR 555.5556 (>= 4.00). 24.001 5100 1500 -900.00V 45U 2W #! Test Tol 45, Sys Tol 0.047, TUR 957.4468 (>= 4.00). 25.001 HEAD {AC Volts @ 400 Hz} 25.002 DISP Set the FUNCTION switch to AC/RMS OR P-P 25.003 DISP Set the PROBE switch to AC OHMS 25.004 DISP NOTE: Read the BOTTOM scale "LOW AC" for all 1.5V range 25.004 DISP ˙ readings. 25.005 5100 1.5 0.800V 0.04U 400H 2W #! Test Tol 0.04, Sys Tol 0.00043, TUR 93.0233 (>= 4.00). 26.001 DISP NOTE: Read the 2nd scale "DC or RMS" for all readings 26.002 DISP ˙ above 1.5V AC. 26.003 5100 5 4.000V 0.25U 400H 2W #! Test Tol 0.25, Sys Tol 0.00245, TUR 102.0408 (>= 4.00). 27.001 5100 15 9.000V 0.5U 400H 2W #! Test Tol 0.5, Sys Tol 0.0042, TUR 119.0476 (>= 4.00). 28.001 5100 50 40.000V 2.50U 400H 2W #! Test Tol 2.5, Sys Tol 0.024, TUR 104.1667 (>= 4.00). 29.001 5100 150 90.00V 5U 400H 2W #! Test Tol 5, Sys Tol 0.0415, TUR 120.4819 (>= 4.00). 30.001 5100 500 400.00V 25U 400H 2W #! Test Tol 25, Sys Tol 0.195, TUR 128.2051 (>= 4.00). 31.001 5100 1500 1000.00V 45U 400H 2W #! Test Tol 45, Sys Tol 0.405, TUR 111.1111 (>= 4.00). 32.001 HEAD {Frequency Response} 32.002 DISP NOTE: Read the BOTTOM scale "LOW AC" for all 32.003 DISP ˙ frequency response measurements. 32.004 RSLT =0.8VAC from 50Hz to 50KHz 32.005 5100 1.5 0.800V 0.04U 50H 2W #! Test Tol 0.04, Sys Tol 0.00043, TUR 93.0233 (>= 4.00). 33.001 5100 1.5 0.800V 0.04U 100H 2W #! Test Tol 0.04, Sys Tol 0.00043, TUR 93.0233 (>= 4.00). 34.001 5100 1.5 0.800V 0.04U 300H 2W #! Test Tol 0.04, Sys Tol 0.00043, TUR 93.0233 (>= 4.00). 35.001 5100 1.5 0.800V 0.04U 3kH 2W #! Test Tol 0.04, Sys Tol 0.00043, TUR 93.0233 (>= 4.00). 36.001 5100 1.5 0.800V 0.04U 10kH 2W #! Test Tol 0.04, Sys Tol 0.00043, TUR 93.0233 (>= 4.00). 37.001 5100 1.5 0.800V 0.04U 30kH 2W #! Test Tol 0.04, Sys Tol 0.00069, TUR 57.9710 (>= 4.00). 38.001 5100 1.5 0.800V 0.04U 50kH 2W #! Test Tol 0.04, Sys Tol 0.00069, TUR 57.9710 (>= 4.00). 39.001 ASK+ X 39.002 ASK- R N P A V 39.003 HEAD {Resistance Calibration} 39.004 HEAD Resistance Meter Zero 39.005 DISP Set the FUNCTION switch to R OHMS 39.006 DISP Set the PROBE switch to AC OHMS 39.007 DISP Select the R x 10 Range. 39.008 DISP Short the probe tips and adjust the "ZERO Adjust" 39.008 DISP knob for a meter reading of 0 ohms. 39.008 DISP Remove the short and adjust the "OHMS adjust" 39.008 DISP knob for a meter reading of infinite ohms. 39.008 DISP 39.008 DISP [27][91]1mREPEAT THE ZERO AND OHMS ADJUSTMENTS UNTIL 39.008 DISP [27][91]1mBOTH READINGS ARE STABLE. 39.009 HEAD {R X 1 Range} 39.010 DISP Select the Rx1 range. # Low Ohms readings taken by HP 3458A 39.011 DISP The HP 3458A will be used in 4-Wire Ohm Mode to 39.011 DISP measure the actual resistance of the 5100B. 39.012 DISP [32]Connect the 5100B to the 3458A as follows: 39.012 DISP 39.012 DISP [32][27][91]1m 5100B TO 3458A 39.012 DISP [32] ŚÄÄÄÄÄÄ> SENSE HI 39.012 DISP [32] OUTPUT HI ÄÄÄÄĮÄÄÄÄÄÄ> INPUT HI 39.012 DISP [32] OUTPUT LO ÄÄÄÄĀÄÄÄÄÄÄ> INPUT LO 39.012 DISP [32] ĄÄÄÄÄÄÄ> SENSE LO 39.013 RESF 10.0Z S 2W #! System Uncertainty not calculated because tolerance or floor unspecified. 39.014 3458 10.0Z E N 4W 39.015 DISP [32]Disconnect the 5100B from the 3458A and 39.015 DISP [32]connect the UUT to the 5100B as follows: 39.015 DISP 39.015 DISP [32][27][91]1m 5100B TO VOM 39.015 DISP [32] OUTPUT HI ÄÄÄÄÄÄÄÄÄÄÄ> Probe Tip 39.015 DISP [32] OUTPUT LO ÄÄÄÄÄÄÄÄÄÄÄ> Ground Clip 39.016 MEME 39.017 MEMI Use keyboard to enter actual UUT reading in Ohms: 39.018 MEME 39.019 MEMC ˙ź 10% #! Test Tol 1, Sys Tol 0.0002, TUR 5000.0000 (>= 4.00). 40.001 HEAD {R x 10 Range} 40.002 DISP Select the R x 10 range. 40.003 DISP [32]Connect the 5100B and 3458A as follows: 40.003 DISP 40.003 DISP [32][27][91]1m 5100B TO 3458A 40.003 DISP [32] ŚÄÄÄÄÄÄ> SENSE HI 40.003 DISP [32] OUTPUT HI ÄÄÄÄĮÄÄÄÄÄÄ> INPUT HI 40.003 DISP [32] OUTPUT LO ÄÄÄÄĀÄÄÄÄÄÄ> INPUT LO 40.003 DISP [32] ĄÄÄÄÄÄÄ> SENSE LO 40.004 RESF 100.0Z S 2W #! System Uncertainty not calculated because tolerance or floor unspecified. 40.005 3458 100.0Z E N 4W 40.006 DISP [32]Disconnect the leads from the 3458A and 40.006 DISP [32]connect them to the UUT as follows: 40.006 DISP 40.006 DISP [32][27][91]1m 5100B TO VOM 40.006 DISP [32] OUTPUT HI ÄÄÄÄÄÄÄÄÄÄÄ> Probe Tip 40.006 DISP [32] OUTPUT LO ÄÄÄÄÄÄÄÄÄÄÄ> Ground Clip 40.007 MEME 40.008 MEMI Use keyboard to enter actual UUT reading in Ohms: 40.009 MEME 40.010 MEMC ˙ź 10% #! Test Tol 10, Sys Tol 0.0015, TUR 6666.6667 (>= 4.00). 41.001 ASK+ R 41.002 ASK- F 41.003 DISP The remaining resistance measurements will be made 41.003 DISP directly from the 5100B. 41.004 HEAD {R x 100 Range} 41.005 DISP Select the R x 100 range. 41.006 RESF 1000Z S 2W #! System Uncertainty not calculated because tolerance or floor unspecified. 41.007 MEMI Enter the actual UUT reading in ohms: 41.008 MEME 41.009 MEMC ź 10% #! T.U.R. not calculated because System Uncertainty not available. 42.001 HEAD { R x 1000 Range} 42.002 DISP Select the R x 1000 range. 42.003 RESF 10000Z S 2W #! System Uncertainty not calculated because tolerance or floor unspecified. 42.004 MEMI Enter the actual UUT reading in ohms: 42.005 MEME 42.006 MEMC ź 10% #! T.U.R. not calculated because System Uncertainty not available. 43.001 HEAD {R x 10K Range} 43.002 DISP Select the R x 10K range. 43.003 RESF 100.0kZ S 2W 43.004 MEMI Enter the actual UUT reading in kohms: 43.005 MEME 43.006 MEMC kź 10% #! Test Tol 10000, Sys Tol 3, TUR 3333.3333 (>= 4.00). 44.001 HEAD {R x 100K Range} 44.002 DISP Select the R x 100K range. 44.003 RESF 1.000MZ S 2W 44.004 MEMI Enter the actual UUT reading in Mohms: 44.005 MEME 44.006 MEMC Mź 10% #! Test Tol 1e+005, Sys Tol 100, TUR 1000.0000 (>= 4.00). 45.001 HEAD {R x 1 MEG Range} 45.002 DISP Select the R x 1MEG range. 45.003 RESF 10.00MZ S 2W 45.004 MEMI Enter the actual UUT reading in Mohms: 45.005 MEME 45.006 MEMC Mź 10% #! Test Tol 1e+006, Sys Tol 3000, TUR 333.3333 (>= 4.00). 46.001 HEAD 46.002 DISP ŚÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄæ 46.002 DISP ³ UUT calibrated to manufacturers specification except ³ 46.002 DISP ³ for the following: ³ 46.002 DISP ³ Calibrated DC/AC voltage and Ohms functions ONLY ³ 46.002 DISP ³ Ohms accuracy ń10% of nominal. ³ 46.002 DISP ³ USE ORANGE CALIBRATION LABEL!!! ³ 46.002 DISP ĄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄŁ 46.003 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 2 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.